Specification
A pipette (micropipette) is an instrument that uses a piston to expand and contract through a spring to achieve liquid aspiration and dispensing. The working principle of a pipette is to use a piston to push out part of the air, then use atmospheric pressure to inhale liquid, and then the piston pushes the air out of the liquid. Pipettes can transfer liquids quantitatively and are widely used in chemistry, biology and other fields.
Features
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The low part is available for autoclaving
Single-channel Adjustable Volume Pipettes |
||||||
Volume Range |
Increment |
Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||
|
|
|
% |
μl |
% |
μl |
0.1-2.5μl |
0.05μl |
2.5 |
2.50% |
0.0625 |
2.00% |
0.05 |
1.25 |
3.00% |
0.0375 |
3.00% |
0.0375 |
||
0.25 |
12.00% |
0.03 |
6.00% |
0.015 |
||
0.5-10μl |
0.1μl |
10 |
1.00% |
0.1 |
0.80% |
0.08 |
5 |
1.50% |
0.075 |
1.50% |
0.075 |
||
1 |
2.50% |
0.025 |
1.50% |
0.015 |
||
2-20μl |
0.5μl |
20 |
0.90% |
0.18 |
0.40% |
0.08 |
10 |
1.20% |
0.12 |
1.00% |
0.1 |
||
2 |
3.00% |
0.06 |
2.00% |
0.04 |
||
5-50μl |
0.5μl |
50 |
0.60% |
0.3 |
0.30% |
0.15 |
25 |
0.90% |
0.225 |
0.60% |
0.15 |
||
5 |
2.00% |
0.1 |
2.00% |
0.1 |
||
10-100μl |
1μl |
100 |
0.80% |
0.8 |
0.15% |
0.15 |
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
10 |
3.00% |
0.3 |
1.50% |
0.15 |
||
20-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
20 |
3.00% |
0.6 |
1.00% |
0.2 |
||
50-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
100-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
100 |
2.00% |
2 |
0.70% |
0.7 |
||
200-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
200 |
0.90% |
1.8 |
0.30% |
0.6 |
||
1000-5000μl |
50μl |
5000 |
0.50% |
25 |
0.15% |
7.5 |
2500 |
0.60% |
15 |
0.30% |
7.5 |
||
1000 |
0.70% |
7 |
0.30% |
3 |
||
2-10ml |
0.1ml |
10ml |
0.60% |
60 |
0.20% |
20 |
5ml |
1.20% |
60 |
0.30% |
15 |
||
2ml |
3.00% |
60 |
0.60% |
12 |
8-channel Adjustable Volume Pipettes |
||||||||
Volume Range |
Increment |
Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||||
|
|
|
% |
μl |
% |
μl |
||
0.5-10μl |
0.1μl |
10 |
1.50% |
0.15 |
1.50% |
0.15 |
||
5 |
2.50% |
0.125 |
2.50% |
0.125 |
||||
1 |
4.00% |
0.04 |
4.00% |
0.04 |
||||
5-50μl |
0.5μl |
50 |
1.00% |
0.5 |
0.50% |
0.25 |
||
25 |
1.50% |
0.375 |
1.00% |
0.25 |
||||
5 |
3.00% |
0.15 |
2.00% |
0.1 |
||||
50-300μl |
5μl |
300 |
0.70% |
2.1 |
0.25% |
0.75 |
||
150 |
1.00% |
1.5 |
0.50% |
0.75 |
||||
50 |
1.50% |
0.75 |
0.80% |
0.4 |
12-channel Adjustable Volume Pipettes |
||||||||
Volume Range |
Increment |
Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||||
|
|
|
% |
μl |
% |
μl |
||
0.5-10μl |
0.1μl |
10 |
1.50% |
0.15 |
1.50% |
0.15 |
||
5 |
2.50% |
0.125 |
2.50% |
0.125 |
||||
1 |
4.00% |
0.04 |
4.00% |
0.04 |
||||
5-50μl |
0.5μl |
50 |
1.00% |
0.5 |
0.50% |
0.25 |
||
25 |
1.50% |
0.375 |
1.00% |
0.25 |
||||
5 |
3.00% |
0.15 |
2.00% |
0.1 |
||||
50-300μl |
5μl |
300 |
0.70% |
2.1 |
0.25% |
0.75 |
||
150 |
1.00% |
1.5 |
0.50% |
0.75 |
||||
50 |
1.50% |
0.75 |
0.80% |
0.4 |
Fixed Volume Pipettes |
||||||||
Volume Range |
Increment |
Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||||
|
|
|
% |
μl |
% |
μl |
||
5μl |
- |
5μl |
1.3% |
0.065 |
1.2% |
0.06 |
||
10μl |
- |
10μl |
0.8% |
0.08 |
0.8% |
0.08 |
||
20μl |
- |
20μl |
0.6% |
0.12 |
0.5% |
0.1 |
||
25μl |
- |
25μl |
0.5% |
0.125 |
0.3% |
0.075 |
||
50μl |
- |
50μl |
0.5% |
0.25 |
0.3% |
0.15 |
||
100μl |
- |
100μl |
0.5% |
0.5 |
0.3% |
0.3 |
||
200μl |
- |
200μl |
0.4% |
0.8 |
0.2% |
0.4 |
||
250μl |
-- |
250μl |
0.4% |
1.0 |
0.2% |
0.5 |
||
500μl |
- |
500μl |
0.3% |
1.5 |
0.2% |
1.0 |
||
1000μl |
- |
1000μl |
0.3% |
3.0 |
0.2% |
2.0 |
||
2000μl |
- |
2000μl |
0.3% |
6.0 |
0.15% |
3.0 |
||
5000μl |
- |
5000μl |
0.3% |
15 |
0.15% |
7.5 |