Biphasic Heart Defibrillator Monitor for Emergency Medical Use

Product Details
Customization: Available
Classification: Physiological Functions of Diagnosis and Monitoring Equipment
Type: Defibrillator Aed
Gold Member Since 2024

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

  • Biphasic Heart Defibrillator Monitor for Emergency Medical Use
  • Biphasic Heart Defibrillator Monitor for Emergency Medical Use
  • Biphasic Heart Defibrillator Monitor for Emergency Medical Use
  • Biphasic Heart Defibrillator Monitor for Emergency Medical Use
  • Biphasic Heart Defibrillator Monitor for Emergency Medical Use
  • Biphasic Heart Defibrillator Monitor for Emergency Medical Use
Find Similar Products
  • Overview
  • Product Paramenters
  • Detailed Photos
  • Company Profile
Overview

Basic Info.

Model NO.
Defi-X
Certification
CE, ISO13485
Group
All
Product Name
Defibrillator Aed
Warranty
1 Year
Waveform
Biphasic
Screen
8 Inch LCD
Mode
Adult, Child
Energy
Delvered
Language
2 5 7 10 20 30 50 70 100 150 200 300 360j
Info
Heart Rate . Lead/Pads. Alarm . on/off . SpO2 . Ae
MOQ
1PCS
OEM
Yes
Battery
Yes
Printer
Yes
Transport Package
Standard Export Packing
Specification
49*54*49mm
Trademark
Mornmed
Origin
Jiangsu, China
Production Capacity
300 Set/Sets Per Month

Product Description

Defibrillator Monitor

 
 
Biphasic Heart Defibrillator Monitor for Emergency Medical Use

Product Paramenters


Biphasic Heart Defibrillator Monitor for Emergency Medical Use
Biphasic Heart Defibrillator Monitor for Emergency Medical Use
Technical Specification: 

Data Acquisition and Analysis
The Defi-X is no longer automatically set to a specific frequency response. There are two diagnostic frequency options: 0.05-150 Hz diagnostic frequency filter and 0.05-40 Hz filtered diagnostic frequency filter 
See chapter 7 for switching between these two options.

Waveform Details 
The following tables show the waveforms provided by the Defi® Xpress when connected to resistive loads of 25, 50, and 100 ohms and set to its maximum output. 
The waveforms are identified by the Defr-X peak current values (Ip) for two-phase operation: 
Duration of the first output phase (tphase1) and duration of the second output phase (tphase2). 
The indicated values are within 10%. 
Waveform: Truncated exponential curve Biphasic 

Output Energy Accuracy 
±10% or 1J(whichever is greater) at 50 ohms
±15% or 1J(whichever is greater) at 25-100 ohms

Charge Time 
< 06 seconds @ 200 J (with a Ni- MH 12 volt  ). 
< 12 seconds @ 360 J (with a Ni- MH 12 volt  after 15 discharges). < 15 seconds @ 360 J (without battery 90% AC mains voltage) 
Indication of the state of charge: acoustic and graphic. 

Output
Adult paddle, pediatric adapter and multipurpose hands-free adapter available. 
Synchronizer: delivers power within 60 msec.
Disarm: Front panel switch. 

Heart Rate Sp02 Alarm 
Auditory: 5-pulse 800 tone with a PW of 150 msec, a PRI of 225 msec, and a repetition interval of 10 seconds. 
Visual: The heart rate alarm flashes the displayed heart rate at 2 Hz. 
This indicator is located at the top of the display and is 0.4" high and 0.28 to 0.840" wide, depending on the number of digits in the heart rate. It is amber, black, white, or the color of the display text. 

Lead Fault Alarm 
Audible: trio tone with 3 pulses, 500 Hz, with a PW of 200 msec and a PRI of 310 msec. 
When the HR alarm is triggered or the pacemaker is on, the lead error tone is repeated at a 20-second repetition interval. 
Visual: When a lead error occurs, the message "LEAD FAULT" is displayed along with a dashed line across the width of the trace. 
The text is 0.18" high and is the color of the display text. 
The dashed line is 5" long and corresponds to the width of the display. The signal is not modulated 

Physiological Alarms (Sp02) (Options) 
Acoustic: See heart rate alarm 
Visual: Physiological alarms make the displayed parameter flash at 2 Hz. 
The color is amber, white, or the same as the colored text on the display. 
Mute Duration: 90 seconds. 

Heart Rate Sp02 Alarm 
Auditory: 5-pulse 800 tone with a PW of 150 msec, a PRI of 225 msec, and a repetition interval of 10 seconds. 
Visual: The heart rate alarm flashes the displayed heart rate at 2 Hz. 
This indicator is located at the top of the display and is 0.4" high and 0.28 to 0.840" wide, 
depending on the number of digits in the heart rate. It is amber, black, white, or the color of the display text. 

Input 
3-wire 5-core patient cable
Size: 8 inch 
Type: color (TFT) 
Screen resolution: 800 x 600 pixels. 
Passing speed: 12.5mm/s25mm/s50mm/s 
Line selection: I II Ill AVR AVL AVF V. 
Frequency response: (user selectable). 
0. 7 to 20 Hz Operating mode
0.3 to 40 Hz Monitoring mode 
0.05 to 150 Hz Analysis mode 
0.5 to 30 Hz AED mode 
Common mode rejection: > =60dB 
Heart rate monitoring: 20 to 300 8PM. 
Heart rate alarms: user selectable. 
Size: 0.125 0.25 0.5 1 2 4 cm/mv and auto ranging. ECG output: 1 V/mv. 
Heart rate response to irregular rhythm: 
(AAMI EC13-1992 Section 4.1.2.1.e.) 

Energy Select
External 2 5 7 10 20 30 50 70 100 150 200 300 360J 

NIBP (Options)
Method: automatic oscillation
Working mode: manual/automatic
Measure: adult: 10-250mmHg
Ement: pediatric: 10-200mmHg
Ement-: neonatal: 10-135mmHg

End-tidal CO2 (EtC02) Options

Pacemaker (Options) 

Frequency Response 
Automatically adjusted to the frequency response of the monitor. 

Battery 
High capacity type: Ni- MH 12V volt - MEDITECH 
Up to 4 hours ECG monitoring or 90 full energy 
charges or 3 hours combined ECG Sp02 and monitoring Proper battery maintenance is required to maintain 
maximum available capacity. 
Low battery indicator: 
Flashing low battery icon on display and flashing 
LED on front panel. 
Recharge speed: 80% in 7.5 hours. 100% in 9 hours. 

General 
Weight: 4.95 kg (basic equipment). 
Size: 13 x 12.5 x 5.3 inches (33 x 31.8 x 13.5 cm). 

Operating Power 
Ni- MH 12 volt - MEDITECH 
SAED Rhythm detection performance 
The algorithm DEFI XPRESS SAED exceeds 
the requirements of ANSI /AAMI. 

Recorder 
Type: high resolution thermal array. 
Annotation: time date ECG lead ECG gain heart rate defibrillation and treatment summary ACLS events. Paper width: 50 mm. 
Paper speed: 25 mm/sec 50 mm/sec 12 leads 50 mm/sec Delay: 6 sec. 



 

Detailed Photos


Biphasic Heart Defibrillator Monitor for Emergency Medical UseBiphasic Heart Defibrillator Monitor for Emergency Medical UseBiphasic Heart Defibrillator Monitor for Emergency Medical UseBiphasic Heart Defibrillator Monitor for Emergency Medical UseBiphasic Heart Defibrillator Monitor for Emergency Medical UseBiphasic Heart Defibrillator Monitor for Emergency Medical Use
 

Company Profile

 
Biphasic Heart Defibrillator Monitor for Emergency Medical UseBiphasic Heart Defibrillator Monitor for Emergency Medical Use
 
Biphasic Heart Defibrillator Monitor for Emergency Medical Use

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier