Product feature:
* Servo Control Mode: air temperature, skintemperature, display humidity;
* Parameter Displaying: set temperature,air temperature, skin
temperature, internal humidity, heat power rate;
* Audible & Visual Alarms: powerfailure, sensor failure,
higher & lower temp, air & skin temperature override,
fan failure, system failure;
* >37ºC Preset Mode: power failure reset memory, alarm silence,
keypad lock, self-testfunction;
* Hood with dual slope; 6 hand-ports;
* RS-232 Output Connector (use for networksystem);
* IV poles;
* Protective guard rail;
* Storage cabinets;
* Can option humidity control.
Specifications of hood
Mattress to Hood: 450 mm
Observation side of hood: Two incline
Operate port: 6 (2 Iris ports)
Tubing grommets: 4
Front access panel: 826×310 mm
Mattress Tray Size: 630×360 mm
Infant bed tilt adjusting ±10° step-less
Technical Specifications
Power requirement ≤400VA
Temperature control mode:
Air and skin servo control
Air Temp Control Range: 20.0ºC-37.0ºC
Temperature override mode to: 37.1ºC-39.0ºC
Skin Temp Control Range: 20.0ºC-37.0ºC
Temperature override mode to: 37.1ºC-38.0ºC
Skin temp sensor accuracy: ≤0.3ºC
Humidity display range: 5%-99%
Temperature fluctuation: ≤0.5ºC
Mattress temperature uniformity: ≤0.8ºC
Temperature rise time: ≤45 min
Noise level within hood environment: ≤55dB
Alarms
Power failure alarm
Fan failure alarm
Sensor failure alarm
High air temperature alarm: +3.0ºC
Low air temperature alarm: -3.0ºC
High skin temperature alarm: +1.0ºC
Low skin temperature alarm: -1.0ºC
Over air temperature alarm: ≤38.0ºC; ≤40.0ºC
Over skin temperature alarm: ≤38.0ºC; ≤39.0ºC
System failure alarm
Alarm silence function
Operational condition
Environment temperature: +20ºC-+30ºC
Environment relative humidity: 30%-75%
Environment air velocity of flow: <0.3meter/s
Others
Mattress tray Max. load 5kg
Transfusion pole Max. load 2kg
Cabinet Max. load 6kg